Humidity Residue Drying Control
Use the material acceptance endpoint to compare equipment configurations; confirm each capability on the selected model.
Moisture and solvent can have different release profiles; first compare small-sample mass loss and residue measurements, then define staged temperature and pressure windows; surface dryness is not an internal-residue result.
Equipment decisions
-
Specify absolute pressure and gauge location; a pump ultimate-pressure rating is not the chamber operating pressure with a wet load.
-
Check sample and tray contact with heated surfaces. Heat transfer changes at low pressure, so the sample core can still be warming after the controller reaches setpoint.
-
Match pump, seals, cold trap or condenser to the released vapour. A standard vacuum-oven configuration does not establish suitability for flammable solvents.
-
Determine completion from material endpoints such as residual moisture, solvent residue or constant mass; record sample temperature, operating pressure and moisture regain after venting.
Trial and acceptance
Record the initial sample condition, loading layout, recipe version and raw measurements. Agree the acceptance endpoint before the trial; repeat at challenging load positions and compare the finished material result.
Catalogue comparison
Candidate models in the current catalogue: ESC-VO1000-T6, ESC-VO125-T6, ESC-VO216-T6, ESC-VO27-T6. These are comparison candidates, subject to the process checks above.
The following specifications apply only to the models listed:
- ESC-VO1000-T6: Internal Chamber Dimensions: 1000*1000*1000 mm
- ESC-VO1000-T6: Operating Temperature Range: A:Room temperature+25~200℃;B:Room temperature+35~300℃;C:Room temperature+45~400℃;D:Room temperature+50~500℃;E:Room temperature+60~600℃
- ESC-VO1000-T6: Heating System: Stainless steel high-speed heater
- ESC-VO1000-T6: Control System: Touchscreen programmable control
- ESC-VO125-T6: Internal Chamber Dimensions: 500*500*500 mm
- ESC-VO125-T6: Operating Temperature Range: A:Room temperature+25~200℃;B:Room temperature+35~300℃;C:Room temperature+45~400℃;D:Room temperature+50~500℃;E:Room temperature+60~600℃
- ESC-VO125-T6: Heating System: Stainless steel high-speed heater
- ESC-VO125-T6: Control System: Touchscreen programmable control
- ESC-VO216-T6: Internal Chamber Dimensions: 600*600*600 mm
- ESC-VO216-T6: Operating Temperature Range: A:Room temperature+25~200℃;B:Room temperature+35~300℃;C:Room temperature+45~400℃;D:Room temperature+50~500℃;E:Room temperature+60~600℃
- ESC-VO216-T6: Heating System: Stainless steel high-speed heater
- ESC-VO216-T6: Control System: Touchscreen programmable control
- ESC-VO27-T6: Internal Chamber Dimensions: 300*300*300 mm
- ESC-VO27-T6: Operating Temperature Range: A:Room temperature+25~200℃;B:Room temperature+35~300℃;C:Room temperature+45~400℃;D:Room temperature+50~500℃;E:Room temperature+60~600℃
- ESC-VO27-T6: Heating System: Stainless steel high-speed heater
- ESC-VO27-T6: Control System: Touchscreen programmable control